Ample Technology ATC – 6900 GC 同樣的品質,更多的選擇。
ATC 6900 GC是美國安普科技中心**推出的氣相色譜儀,該儀器以強大的APC技術(Advanced pneumatic control先進氣壓控制)全面控制整體系統的流量;創新的指尖觸摸直接操作直觀可感知的操作界面;強大的網絡(LAN)通訊加速了數據處理速度。
國際通用的強大ATC-Clarity色譜數據處理及儀器控制軟件。該軟件符合21CFR Part 11.
ATC 6900 GC色譜主機與Tekmar公司的StratumPTCTM吹掃捕集濃縮儀、AquatekTM100液體自動進樣器、AtomxTM固/液全自動吹掃捕集儀、VersaTM頂空自動進樣器、HT3TM靜態/動態頂空自動進樣器等前處理設備,可**、輕松地完成您的分析工作。
A. Features(儀器特征)
觸摸顯示直觀圖形界面
Intuitive graphic user interface with large touchpad color LCD (5.7”).
強大的APC使分析更精確
Strongly enhanced powerful APC for accurate and precise results
自動存儲20種分析方法
Automatic loading and saving up to 20 different analytical methods.
高速的網絡數據傳輸
High speed data process by network (LAN) communication.
可安裝多種進樣口和檢測器
Various inlets and detectors installable up to three per each.
自動關機
Time control : Automatic turn-on/off
10個加熱區
Ten heated zone :
1 heated oven zone
3 heated injector zone
3 heated detector zone
2 heated valve zone
1 heated methanizer zone
B. Specifications(規格詳述)
1. Gas Chromatograph ATC 6900 GC Oven System Module(主機系統)
柱箱容積:14L
Usable volume : 14L
自動控制冷卻控制
Automatic cooling under processor control
溫度控制范圍:環境溫度4-450℃
Temperature operating range : 4℃ above ambient to 450℃ or more
25平臺,26階程序升溫
Programming: 25 ramps, 26 steps temperature program
**升溫速率:100℃/min
Maximum set-point heating rate: 100℃/min
冷卻速率:450℃ 降至50℃ 需6.5分鐘
Cool down rate: 6.5 minutes from 450℃ to 50℃
控溫精度:±0.01℃ (等度), ±0.1℃ (梯度)
Temperature stability : ±0.01℃ (Isothermal), ±0.1℃ (Gradient)
2. Injector ( Maximum no. of inlet installation: three)(注樣器系統,可安裝3個)
(1) Packed inlet(填充柱進樣系統)
**設定溫度:450
Maximum Temperature : 450 oC
壓力設定范圍:
Pressure setting range : 0.01 ~ 100 psi
總流量設定范圍:
Total flow setting range : 0.1~100 ml/min
(2) Capillary inlet (Split/Splitless)(毛細管進樣系統(分流/不分流))
**設定溫度
Maximum Temperature : 400oC
壓力設定范圍:
Pressure setting range : 0.01 ~ 100 psi
總流量設定范圍:
Total flow setting range :
0.1 ~ 400 ml/min N2
0 ~ 1000ml/min He
3. Advanced Pneumatic Control(氣壓控制)
6個APC模塊控制注樣器、檢測器和輔助氣
所有氣體均有APC控制
All gas flows controlled by APC
程序流量:5階
Programmed Flow : 5step
程序壓力:5階
Programmed Pressure : 5step
檢漏測試
Leak detection
檢漏報警后,自動運行到關閉狀態
After an alarm for a shortage of gas, it automatically saves the method to method no. 0 and shuts down
4. Detectors (檢測器)
- Maximum no. of detector installation: three(可安裝3個檢測器)
- Data Acquisition Rate : 200 Hz (采樣頻率)
(1) Flame Ionization Detector(FID)
(2) Thermal Conductivity Detector(TCD)
MDL : 2.5 ng/ml (Standard), 400 pg/ml (uTCD)
(3) Nitrogen-Phosphorous Detector(NPD)
(4) Flame Photometric Detector(FPD)
(5) Pulsed Discharge Detector(PDD)
Helium ionization mode(PDHID) :MDL (organic compound : low ppb;permanent gas : low ppm)
Electron capture mode(PDECD) : MDL – 10-15
(6) Electron Capture Detector (ECD